Getinge GE 672x920x1350-2, Autoclave sterilization AB
EUROPE (Western and Northern)
Steam (non-process): 9 kg; peak/min. 0.9 kg; 2.7-3.0 bar
Steam process: Use 15 kg; peak/min. 0.9 kg; 1.5-2.0 bar
Water pump: use 30 liters; peak/min. 10 liters; 4.0-6.0 bar
Cooling water: 280 liters; peak/min. 32 liters; 2.0-3.0 bar
Air: usage 0.4 Nm3; peak / min. 0,2 Nm3; 6.0-7.0 bar
Air process: usage 3.4 Nm3; peak/min. 0,4 Nm3; 4.0-5.0 bar
Cooling water return: 280 liters; peak/min. 32 litres
Drainage: peak / min. 20
Three-phase with voltage 400 volts; 4,3 kW
Frequency: 50Hz
Recommended fuse: 20 Ampere
Rated power: 3 Kw
Rated current: 7A
Short circuit current: 6 kA
Maximum single load: 6 A
Heat emission:
Load area: 0.5 Kw
Discharge area: 0.5 Kw
Service Area: 2.6 Kw
Sound level: < 70 dBa