Getinge GE 700X1450X1350-1, Autoclaves/Sterilizing Ovens/Freeze Dryers
EUROPE (Western and Northern)
The Autoclave is supplied with the following process cycles and test facilities:
1. Vacuum Cycle
2. Leak Test
3. WarmUp and ALRT combined
Mechanical Enclosure
Fascia Panels
the sterilizer is a recessed installation with fascia panels to the loading side only.
Material: Stainless steel 1.4301 (AISI 304).
Front surface texture: Tetra polished Ra< 1,2 micrometer
Thickness 1.0 mm.
Loading area / Plantroom area Barrier
There is no requirement for a barrier.
Pressure Vessel
Loading Height: 300 mm from floor to the chamber bottom.
Chamber
Internal Dimensions
Useable width 700 mm
Useable height 1450 mm
Useable depth 1350 mm
Construction:
Cross section: Rectangular
Material: AISI 316L acid proof stainless steel
Design pressure: 3 bar (g) & full vacuum
Design temperature: 143 °C
Welds: Robotic or semi-automatic whenever possible
Internal Chamber and Ports Finish
internal surface ground and polished to a high lustre
Internal Ra value 0.6 microns
1 Door
Design: Fully automatic, horizontal sliding
Opening and closing: Automatic button controlled
Door operation: Pneumatic motor
Door safety: Stops and reverses if obstructed. Door close isolation keyswitch to prevent door closing if access to chamber is required.
Internal door surface: AISI 316L stainless steel
Door pins: AISI 316 stainless steel
Media (Utility) Requirements
Please refer to the installation drawing for details of sizes & consumptions.
Requirements:
Steam
Compressed Air Supply
Potable Water Supply
Cooled Water Supply
Condense