GE Healthcare Avance CS2
EUROPE (Western and Northern)
• 15-inch touchscreen with CARESCAPETM inspired user interface
for a unified experience on the Carestation.
Features
Various controlled ventilation modes: volumetric ventilation,
pressure ventilation, etc.
Intuitive user interface with touchscreen for easy programming
Flexibility to adapt to various clinical situations and
patient types
Advanced safety features, including alarms for
abnormal flow or pressure
Integration with medical information systems for
efficient clinical data management
Technical Details
Dimensions : 139 (H) x 77 (W) x 76 (D)
Weight : 147 kg
Physical Specifications
Dimensions
Height: 139 cm
Width: 77 cm
Depth: 76 cm
Weight: 147 kg1
Top shelf
Weight limit: 34 kg/75 lb
Width: 69.7 cm
Depth: 44 cm
Work surface
Height: 81.7 cm/32.2 in
Size: 2640 cm²/409 in²
Upper left GCX/DO dovetail
GCX length: 13.2 cm
DO dovetail length: 23.2 cm
Upper right DO dovetail
Length: 34 cm
Drawers (internal dimensions)
Height: 17.5 cm/6.9 in
Width: 33 cm/13 in
Depth: 26.5 cm/10.4 in
Absorber bag arm (optional)
Arm length: 39.8 cm/15.7 in
Bag arm height (adjustable):
87 cm/34.3 in
113 cm/44.4 in
Casters
Diameter: 13 cm
Brakes: Central Brake
Lower right GCX rail
Length: 41 cm
Ventilator Operating Specifications
Modes of ventilation - standard
VCV (Volume Control) Mode with tidal volume compensation
Modes of ventilation - optional
Pressure Control
Pressure Controlled Ventilation-Volume Guaranteed (PCV-VG)
SIMV (Synchronized Intermittent Mandatory Ventilation)
(volume and pressure)
PSVPro* (Pressure Support with Apnea backup)
CPAP+PSV (Pressure support mode)
SIMV PCV-VG
Ventilator parameter ranges
Tidal volume range: 5 to 1500ml (Volume Control,
PCV-VG and SIMV volume 20 to 1500ml)
(PCV modes 5 to 1500ml)
Incremental settings:
20 to 50 mL (increments of 1 mL)
50 to 100 mL (increments of 5 mL)
100 to 300 mL (increments of 10 mL)
300 to 1000 mL (increments of 25 mL)
1000 to 1500 mL (increments of 50 mL)
Minute volume range: Less than 0.1 to 99.9 L/min
Pressure (Pinspired) range: 5 to 60 cmH2O
(increments of 1 cmH2O) 5 to 1500 mL volume delivery
Pressure (Pmax) range: 12 to 100 cmH2O
(increments of 1 cmH2O)
Pressure (Pinspired) range: 5 to 60 cmH2O
(increments of 1 cmH2O) 5 to 1500 mL volume delivery
Pressure (Pmax) range: 12 to 100 cmH2O
(increments of 1 cmH2O)
Pressure (Psupport) range: Off, 2 to 40 cmH2O
(increments of 1 cmH₂O)
Rate:
4 to 100 breaths per minute for Volume Control and
Pressure Control; 2 to 60 breaths per minute for SIMV,
PSVPro and SIMV PCV-VG; 4 to 60 bpm for
CPAP+PSV (increments of 1 breath per minute)
Inspiratory/ expiratory ratio: 2:1 to 1:8 (increments of 0.5)
Inspiratory time: 0.2 to 5.0 seconds (increments of 0.1 seconds)
(SIMV, PSVPro and CPAP PSV)
Trigger window: 0 to 80% (increments of 5%)
Flow trigger: 1 to 10 L/min (increments of 0.5 L/min)
0.2 to 1 L/min (increments of 0.2 L/min)
Inspiration termination level: 5 to 75% (increments of 5%)
Inspiratory Pause range: 0-60%
Positive End Expiratory Pressure (PEEP)
Type: Integrated, electronically controlled
Range: OFF, 4 to 30 cmH2O (increments of 1 cmH2O)
Ventilator performance
Pressure range at inlet: 240 kPa to 700 kPa
(35 psig to 100 psig)
Peak gas flow: 120 L/min + fresh gas flow
Flow valve range: 1 is 120 L/min
Flow compensation range: 150 mL/min to 15 L/min
Ventilator Accuracy
Delivery/monitoring accuracy
Volume delivery:
> 210 mL = better than 7%
≤ 210 mL = better than 15 mL
< 60 mL = better than 10 mL
Pressure delivery: +10% or ±3 cmH2O
PEEP delivery: ±1.5cmH2O
Volume monitoring:
> 210 mL = better than 9%
< 210 mL = better than 18 mL
< 60 mL = better than 10 mL
Pressure monitoring: ±5% or ±2 cmH2O
Alarm settings
Tidal volume (VTE):
Low: OFF, 1 to 1500 mL
High: 20 to 1600 mL, OFF
Minute volume (VE):
Low: OFF, 0.1 to 10 L/min
High: 0.5 to 30 L/min, OFF
Inspired oxygen (FiO2):
Low: 18 to 99%
High: 19 to 100%, OFF
Apnea alarm:
Mechanical ventilation ON:
< 5 mL breath measured in 30 seconds
Mechanical ventilation OFF:
< 5 mL breath measured in 30 seconds
Low airway pressure:
4 cmH2O above PEEP
High pressure:
12 to 100 cmH2O (increments of 1 cmH2O)
Sustained airway pressure:
Mechanical ventilation ON:
Pmax < 30 cmH2O, the sustained limit is 6 cmH2O
Pmax 30 to 60 cmH2O, the sustained limit is 20% of Pmax
Ртак > 60 cmH2O, the ma sustained limit is 12 cmH2O
PEEP and mechanical ventilation ON
Sustained limit increases by PEEP minus 2 cmH2O
Mechanical ventilation OFF:
Pmax ≤ 60 cmH2O, the sustained limit is 50% of
Pmax Pmax > 60 cmH2O, the sustained limit is 30 cmH2O
Subatmospheric pressure: Paw < -10 cmH2O
Audio pause countdown clock: 120 to 0 seconds
Ventilator Components
Flow transducer
Type: Variable orifice flow sensor
Dimensions: 22 mm OD and 15 mm ID
Location: Inspiratory outlet and expiratory inlet
(Optional autoclavable sensor available)
Oxygen sensor
Type: Optional galvanic fuel cell or paramagnetic with
Airway Module option