Aisys CS 2 Specifications:
General
Brand: GE
Category: Anaesthetic Equipment
Dimensions
Overall: 52.7 in (133.9 cm) H x 28.92 in (73.45 cm) W x 34.8 in (88.4 cm) D
Top Shelf: 21.66 in (54.8 cm) W x 17.5 in (44.45 cm) D
Work Surface: 365.5 in² (2358.1 cm²)
Drawers (small): 4.13 in (10.5 cm) H x 14.88 in (37.8 cm) W x 14.82 in (37.64 cm) D
Casters Diameter: 5 in (12.5 cm)
Overall Weight: 370 lbs (168 kg)
Top Shelf Weight Limit: 100 lbs (46 kg)
Work Surface Height: 34.71 in (88.17 cm)
Drawers (large): 5.91 in (15 cm) H x 14.88 in (37.8 cm) W x 14.82 in (37.64 cm)
Aisys Ventilation Modes
Standard: Volume control mode with
tidal volume compensation
Optional: Pressure Control and PCV-VG, SIMV, PSVPro
Notification of spontaneous breathing:
Patient-generated breaths will change pressure and
flow waveform color for immediate clinician notification
7900 Smartvent Parameters
Tidal volume range: 20 to 1500 mL
(Volume control and SIMV modes)
Incremental settings: 20 to 50 mL (increments of 1 mL),
50 to 100 mL (increments of 5 mL), 100 to 300 mL
(increments of 10 mL), 30 to 1000 mL (increments for 25 mL),
1000 to 1500 mL (increments of 50 mL)
Minute volume range: 0 to 99.9 L/min
Pressure (P) Inspired range: 5 to 60 cm H2O
(increments of 1 cm H2O),
5 to 1500 mL volume delivery
Pressure (P) max range: 12 to 100 cm H2O
(increments of 1 cm H2O)
Pressure (P) support range: Off, 2 to 40 cm H2O
(increments of 1 cm H2O)
Rate: 4 to 100 breaths per minute for Volume Control and
Pressure Control; 2 to 60 breaths per minute for SIMV,
PSVPro and SIMV-PC+PSV (increments of 1 breath per minute)
Inspiratory/expiratory ratio: 2:1 to 1:8 (increments of 0.5)
Inspiratory time: 0.2 to 5.0 seconds (increments of 0.1 seconds)
(SIMV and PSVPro)
Trigger window: 0 to 80% (increments of 5%)
Flow trigger: 1 to 10 L/min (increments of 0.5 L/min),
0.2 to 1 L/min (increments of 0.2 L/min)
Inspiration termination level: 5 to 75% (increments of 5%) -
Rise Rate 1-10 (PCV, PCV-VG, PSV, SIMV and PCPro)
Features:
eco flow provides information that may help clinicians
maintain their desired inspired oxygen concentration.
During low and minimum flow anesthesia, ecoFLOW displays
agent consumption data that can help mitigate wasteful
overdelivery of fresh gas flow.
Using fewer agents means spending less money, which can
positively impact your bottom line and the environment when
waste gases are reduced.