Menu

Agilent 1260 Infinity ELSD HPLC Evaporative Light Scattering Detector

Ref : 2625435-18-CP
Condition : Used
Manufacturer : Agilent
Model : 1260 Infinity ELSD
Short Description : HPLC Evaporative Light Scattering Detector
Year(s) : -
Quantity : 1
Location : Seller or machines location:
AMERICA North (USA-Canada-Mexico)

The Agilent 1260 Infinity Evaporative Light Scattering Detector is designed to detect compounds
in the eluent from high performance liquid chromatography (HPLC), micro-HPLC,
gel permeation chromatography (GPC) or counter current chromatography (CCC).
It is capable of monitoring eluent flow rates from 5 µl/min to 5 ml/min.
Evaporative light scattering detection is a universal technique which can detect any non-volatile analyte.
Detection does not depend on the absorption of radiation and is not affected by the absorption characteristics of the solvent;
thus solvents which absorb UV radiation can be used.

Specifications ELSD:
Detection: High Sensitivity Photomultiplier
Light Source: Selected High Efficiency Blue LED
Temperature Range: Ambient to 100°C
Gas Flow Control: Manual and computer controlled nebulization gas flow and auxiliary gas flow.
Gas Consumption: Less than 4 l/min for the Large Flow Nebulizer, less than 3 l/min for all other nebulizers.
Gas pressure: 3.5 bar (51 psi) typical, min. 3 bar (44 psi) to max. 4.5 bar (67 psi)
Eluent Flow Rate
Micro Flow Nebulizer: 5 µl/min to 50 µl/min
Semi Micro Flow Nebulizer: 40 µl/min to 1 ml/min
Standard Flow Nebulizer: 500 µl/min to 2 ml/min
Large Flow Nebulizer: 1.5 to 5 ml/min
Rapid Resolution Nebulizer: 0.2 to 1.4 ml/min

Instrument Control: Microprocessor with stand alone manual keypad or Windows based PC control (see specifications for your instrument control software)
Operating Parameters Control: Liquid Crystal Digital Panel
Signal Drift: Less than 1 mV/h
Signal Noise: < ± 1 mV
Signal Output: 0-1000 mV (Analog), RS-232 (Digital)
Inputs: Remote Autozero (Contact Closure or TTL Signal), Remote Powerdown Mode (Contact Closure or TTL Signal)
Power Down Mode: General, Standby, Cleaning
Zero Control: Manual Auto Zero and Remote Auto Zero
Interface: RS-232 I/O Serial Output
Power: 100 V / 2.1 A, 115 V/60 Hz, 1.8 A or 230V/50 Hz, 1.7 A
Dimensions: 250 mm (10″) W x 450 mm (18″) H x 550mm (22″) D
Weight: 18.5 kg (40 lb.)