SP Scientific Genevac Rocket Synergy 2 RKS2-12060-SPW
AMERICA North (USA-Canada-Mexico)
Manufacturers Specifications:
Maximum rotor speed: 1760 rpm (ask rotor) or 1530 rpm (bowl rotor)
Flask rotor capacity: 6 x 400 ml
Bowl rotor capacity: 5 litres
Nominal sample load: g force 500 g
Temperature control range: 30 ºC to 60 ºC
Vacuum Pump: (Integral)
Type: Diaphragm pump
Ultimate system vacuum < 3mbar
Condenser (Integral)
Type: Aluminium water jacket
Minimum coolant temp: -15 ºC
Required coolant low rate: 1.5 ± 0.5 litres/minute
Max. coolant inlet pressure: 30 psi (2 bar g)
Inert Gas Supply (IGP option only)
Maximum pressure 1.5 bar g (2.5 bar abs.)
Minimum pressure 1 bar g (2 bar abs.)
Flow rate (nominal) 30 litres/minute
Hose length 2.5 m
Connector type 3/8” BSP female
Emissions
Noise (@ 1 metre) 63 dB(A)
Operational Environment
Ambient temperature: 15 ºC to 30 ºC
Relative humidity: 10-80% non-condensing
Altitude Sea-level: to 1600 m
Min. ventilation air-gap: 75 mm
Ingress protection rating: IP 20
Installation environment: Indoor only. Static-dissipative laboratory or similar
WEIGHT 205 lbs
DIMENSIONS 30 × 32 × 28 in
INCLUDES Power Cord
VOLTAGE 120V
HERTZ 60Hz