ZEISS Cirrus Photo 800
AMERICA North (USA-Canada-Mexico)
Features:
Broader Clinical Insights
Comprehensive
Specifications:
Field angle 45° and 30°
Pupil diameter ≥ 4.0 mm; ≥ 3.3 mm (30° small pupil mode)
≥ 2.0 mm for OCT scans only
Refractive error compensation +35 D … -35 D, continuous
Working distance 40 mm (patient’s eye – front lens)
Fixation targets External and internal
Internal Attention mode and free position or programmed sequences
Monitor 23” TFT (1920 x 1200)
Instrument table Asymmetric, suitable for wheelchairs
Database Patient information and images with field angle, FA time, R/L recognition and date of visit are stored
Accessories Network printer, sliding keyboard shelf, network isolator,
FORUM eye care data management system
Fundus camera
Capture modes
Color, red-free, blue, red and fundus autofluorescence pictures, as well as pictures of the anterior segment, fluorescein angiography and ICG angiography
Filters
Filters for green
blue and fundus autofluorescence images
UV/IR barrier filters
+ FA + ICGA: exciter and barrier filters
Capture sequence From 1.5 seconds (depends on flash energy)
Capture sensor CCD 5.0 megapixels
Xenon flash lamp 16 flash levels (30 Ws)-24 flash levels (80 Ws)
OCT
Technology Spectral domain OCT
Optical source Superluminescent diode (SLD), 840 nm
Scan speed 27,000 A-scan per second
A-scan depth 2.0 mm (in tissue), 1024 points
Resolution Axial 5 μm (in tissue), transverse 15 μm (in tissue)
Computer
Operating system Windows Embedded Standard 7
Hard drive
Storage of over 30,000 fundus images with OCT cube scans (present size: 320 GB)
Interfaces USB ports and network connectors, DVI port
Export/import
Image formats: BMP, TIFF, JPEG, PNG
Patient list, DICOM MWL, DICOM storage
Dimensions
Main unit (W 16.1 x D 18.9 x H 26.8 inches)
Weight (main unit) (72.7 lbs)
Rated voltage 100 … 240 V ±10%
Frequency 50 / 60 Hz
Power consumption 400 VA (w/o instrument table)