Welch Allyn Spot LXi Vital Signs Monitor
AMERICA North (USA-Canada-Mexico)
Certified Refurbished
Manual weight, height, respiration rate, and pain level entry.
SureBP standard on all configurations.
Compatible with some select digital scales.
Calculates body mass index.
50-reading memory.
Connectivity-ready.
Fast, accurate, and flexible.
Extremely easy to use.
Specifications
Non-invasive Blood Pressure (NIBP)
Cuff Pressure Range: 0 to 300 mmHg (0 to 40.0 kPa).
Systolic Range: 60 to 250 mmHg (8.0 to 33.3 kPa).
Diastolic Range: 30 to 160 mmHg (4.0 to 21.3 kPa).
Mean Arterial Pressure Range: 40 to 190 mmHg (5.3 to 25.3 kPa).
Blood Pressure Accuracy: Meets or exceeds AAMI SP10:2002.
Overpressure Cutoff: 315 mmHg ± 15 mmHg (42.0 kPa ± 1.9 kPa)
Blood Pressure Determination Time: Typically 15 seconds.
Pulse Rate Range: 35 to 199 bpm.
Pulse Rate Accuracy ± 5.0%.
Temperature
Temperature Range:
SureTemp Plus: 26.7 to 43.3 °C.
Braun ThermoScan PRO 4000: 20 to 42.2 °C.
Calibration Accuracy:
SureTemp Plus: ± 0.1 °C (Monitor Mode).
Braun ThermoScan PRO 4000: ± 0.2 °C to 35.5 °C to 42 °C ± 0.25 °C (outside this temperature range).
Display Resolution: -17.7 °C.
SureTemp Plus Predict Time
Oral: Approx. 4 to 6 seconds.
Adult Axillary: Approx. 12 to 15 seconds (age 18 years and older).
Paediatric Axillary: Approx. 10 to 13 seconds (age 17 years and younger).
Rectal: Approx. 10 to 13 seconds.
Probe and Probe Well
Oral/Axillary: Blue.
Rectal: Red.
SpO2
Performance Measurement: 1 to 100%.
Range: 25 to 240 bpm.
SpO2 Accuracy: 70 to 100%.
Pulse Rate Accuracy: 25 to 240 bpm ± 3 digits.
Mechanical
Dimensions
Height: 27 cm.
Length/PRO 4000: 20.32 cm.
Length/SureTemp Plus: 19 cm.
Depth: 13.34 cm.
Weight: 3.4 kg.
Electrical
AC Power Supply: 8 V DC at 0.75 A.
Rechargeable sealed lead acid internal battery.
Battery Recharge Time: 90 to 100% after 6 hours; 1 additional hour for PRO 4000;
4 additional hours for external printer.
Simultaneously operates device and charges battery when AC power adapter
is plugged in.
Wireless Radio
Wireless Network Interface: IEEE 802.11b DSSS, WiFi compliant.
Frequency: 2.4 to 2.4835 GHz.
Environmental
Operating Temperature (Device): 10 C to 40 degrees C.
Operating Relative Humidity: 15 to 95% (non-condensing).
Operating Altitude: -170 to +4,877 m.
Storage Temperature
Device with SureTemp Plus: -25 to 55 degrees C.
Device with PRO 4000: -20 to 50 degrees C.
Storage Humidity: 15 to 95% (non-condensing).
EMC Standard: IEC/EN 60601-1-2:2001.