Specifications:
*Specifications are based on nominal voltages of 115V in ambients of 22°C to 25°C.
Temperature
Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .±0.1°C
Range . . . . . . . . . . . . . .+5°C above ambient to +55°C (131°F)
Uniformity . . . . . . . . . . . . . . . . . . . . . . . . . . .±0.3°C @ +37°C
Low Temp Tracking Alarm . . . . Fixed at 1.0°C below setpoint
Temperature Safety
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . .Precision thermistor
Setability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0.1°C

CO2
CO2 Control . . . . . . . . . . . . . . . . . . . . . . . .Better than ±0.1%
CO 2 Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-20%
Inlet Pressure . . . . . . . . . . . . . .15 PSIG (1.034 bars), ±5 PSIG
CO 2 Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .T/C
Readability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0.1%
Setability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0.1%
Low CO 2 Tracking Alarm . . . . .Fixed at 1.0% below setpoint
High CO 2 Tracking Alarm . . . . .Fixed at 1.0% above setpoint

Humidity
RH . . . . . . . . . . . . . . . . . .Ambient to 95% @ +37°C (98.6°F)
Humidity Pan . . . . . . . . . . . . . . . . . .0.8 gal. (3 liters) standard
Fittings
Fill Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3/8” barbed
Drain Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1/4” barbed
Access Port . . . . . . .1-1/4” (3.18cm) removable neoprene plug
CO2 Inlet . . . . . . . . . . . . . . . . . . . . . . . . . . . .1/4” hose barbed

Unit Heat Load
115V . . . . . . . . . . . . . . . . . . . . . . . . . .344 BTUH (100 Watt)
Shelves
Dimensions . . . . . . . . . . . . .18.5” x 18.5” (47.0cm x 47.0cm)
Construction . . . . . . . . . . . . . . . . . . .Stainless steel, perforated
Surface area . . . . . . . . . . . . . . .2.4 sq. ft. (0.22 sq. m) per shelf
Max. per Chamber . . . . . . . . . . . . . . . .38.4 sq. ft. (3.6 sq. m)
Loading . . . . 35 lbs (16kg) slide in and out, 50 lbs (23kg) stationary
Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Maximum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Construction
Water Jacket Volume . . . . . . . . . . . . . . . .11.7 gal. (43.5 liters)
Interior Volume . . . . . . . . . . . . . . . . . .6.5 cu. ft. (184.1 liters)
Interior . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Stainless steel
Exterior . . . . . . . . . .18 gauge, cold rolled steel, powder coated
Outer Door Gasket . . . . . .Four-sided, molded magnetic vinyl
Inner Door Gasket . .Removable, cleanable, feather-edged, silicone

Electrical
115V . . . . . . . . . . . . .90-125VAC, 50/60 Hz, 1 PH, 3.6 FLA
Circuit Breaker/Power Switch . . . . . . . . . . . . . . .6 Amp/2 Pole
Alarm Contacts . .Power interruption, deviation of temp., CO 2 customer
connections through jack on back of unit. 30V, 1A max.
Installation Category . . . . . . . . . . . . . .Overvoltage Category II
Pollution Degree 2
Maximum Leakage CurrentWith ground disconnected, 0.65mA
Maximum permissible leakage, 3.5mA

Dimensions
Interior . . . . . . . . . . . . . . . . . .21.3” W x 26.8” H x 20.0” F-B
. . . . . . . . . . . . . . . . . . . . . . . . .(54.1cm x 68.1cm x 50.8cm)
Exterior . . . . . . . . . . . . . . . . . .26.3” W x 39.5” H x 25.0” F-B
. . . . . . . . . . . . . . . . . . . . . . . .(66.8cm x 100.3cm x 63.5cm)

Weight (per unit)
Net . . . . . . . . . . . . . . . . . . . . . . .265 lb. (120.2 kg)
Net Operational . . . . . . . . . . . . .365 lb. (165.6 kg)
Shipping . . . . . . . . . . . . . . . . . . .324 lb. (147.0 kg)

Safety Specifications:
Indoor use only
Altitude . . . . . . . . . . . . . . . . . . . . . . . .2000 meters
Temperature . . . . . . . . . . . . . . . . . . . .5°C to 40°C
Humidity . . . . . . . . . . .80% RH at or below 31°C,
. . . . . . . .decreasing linearly to 50% RH at 40°C
Mains Supply Fluctuations . . Operating Voltage Range
Installation Category 21