V1200T Touch Screen
MIDDLE EAST
Brand Axia
Features
The Axia V1200T monitor is an intuitive touch screen patient monitor;
patient information along with vital sign settings can be quickly modified
to meet the needs of a patient’s changing condition. Its upgradable design fits-in
and moves amongst many patient care areas.
Offers ECG with arrhythmia detection, standard measurements of:
non-invasive blood pressure, motion tolerant SpO2, temperature, respiration rate and capnography.
The Axia V1200T monitor offers several connective solutions
to network multiple monitors and/or manage patient data
on an electronic medical records platform or a based hospital information system.
High resolution 12.1 inch touch-screen display
Capnography with EtCO2 monitoring (optional)
Auto setting of alarm limits
Color-coded alarms
Graphical and tabular trends (72 hours)
Patient data entry
OxiCRG, drug dose calculation
Trace: 8 wave forms
Trend Time: 1 – 72 Hours
Networking capability (wireless and non-wireless)
Color of the waveforms can be changed
Battery backup (removable batteries)
Weight: 8.45 lbs.
Applications: Neonatal, pediatric and adult patients
GENERAL SPECIFICATIONS
Display: 12.1 inch color touch screen
Trace: 8 waveforms
Indicator: Alarm indicator, Power indicator, QRS beep and alarm sound
Trend time: 1 – 72 hour
Recorder: Built-in, thermal array
3 channels Record width: 48mm
Recorder paper: 50mm
Record speed: 25mm/s, 50mm/s
Networking
Industry standard 802.11b/g wireless network
Power
Source: External AC power or internal battery
AC Power: 100 ~ 240VAC, 50/60Hz, 150VA
Battery: Built-in & rechargeable lithium ion
Operating Time: 3+ hours
Environmental Specifications
Temperature: Operating: 5 ~ 40 °C, Storage: -20 ~ 65 °C
humidity range: Operating: ≤80 % Storage: ≤80 %
ECG
Input: 5-lead ECG Cable and standard AAMI line for connection
Lead Choice: I, II, III, aVR, aVF, aVL, V, V1-V6, TEST
Gain Choice : x0.5, x1, x2, x4
Frequency Characteristic: 0.05 ~ 35 HZ (+3dB)
ECG Waveforms: 7 channels
Penetration Voltage: 4000VAC 50/60Hz
Sweep Speed: 12.5, 25, 50 and 100 mm/sec (left to right or right to left)
HR Display Range: 30 ~ 300bpm
Accuracy: ±1bpm or ±1%, whichever is greater
Alarm Limit Range Setting: upper limit 100 ~ 200bpm, lower limit 30 ~ 100bpm
RESP
Measure Method: RA-LL impedance
Range: 0 ~ 120 rpm
Accuracy: ±3 rpm
Alarm Limit Setting: upper limit 6 ~ 120 rpm, lower limit 3 ~ 120 rpm
Sweep Speed: 12.5, 25, 50 and 100 mm/sec (left to right or right to left)
Sp02
ASpO2: Anti-motion SpO2
SpO2% Range: 0-100%
SpO2 Accuracy: ±2% (70 ~ 100%, non-motion), ±3% (70 ~ 100%, motion)
Pulse Rate Range: 30-250 bpm
Pulse Rate Accuracy: ±2 bpm (non-motion, ±3 bpm (motion)
Alarm Limit Setting: upper limit 70 ~ 100%, lower limit 70 ~ 100%
SpO2 Probe: Red light LED wavelength 660nm±5nm, Infrared light LED wavelength 940nm±10nm