Tomey AL-100 A-Scan Biometer
ASIA (South East)
Colour touch screen
Automated / manual measurement
Contact / immersion mode
6 IOL calculation formulae
Internal data storage
Memory card drive
Range:
Axial length: 15 ~ 40 mm
ACD velocity: 1.8 ~ 7 mm
Lens thickness: 2 ~ 6 mm
Minimum display unit: 0.01 mm
Accuracy: ± 0.1 mm
IOL calculation:
SPK II, SPK/T, HOLLADAY, Hoffer Q, HAIGIS optimized, HAIGIS standard, SHOWA
Biometry probe
Type: Solid state
Fixation light: Built in the probe, Red LED
Transducer frequency: 10 MHZ ± 10%
Tip radius: Ø 5 mm (concaved)
Dimentions / Weight: Ø 8 x 100 mm / 30g
Data Control
Display: 5.7 inches Color Touch Panel STN LCD
Printer: Thermal printer
Dimensions and Electric Requirement
Dimensions: 220(W) x 222(D) x 275(H) mm
Weight: Approx. 4kg (8.8 lbs.)
Power supply: AC 100 ~ 240 V (adjusts automatically)
Frequancy: 50 / 60 Hz
Consumption power: 35 ~ 42 VA