The Thermo Scientific Thermoflex 900 Recirculating Chiller is an air-cooled unit that uses ambient-temperature room air to remove heat through its condenser.
This chiller offers powerful cooling abilities, a large temperature range, and sophisticated controller technology.
The Thermoflex 900 offers up to 10% greater cooling capacity than comparable units.
With a highly reliable recirculation system and integrated air and fluid filters, this recirculating chiller has a low risk of leaks and low lifetime maintenance cost.
Features of the Thermoflex 900 include auto refill, anti-drainback, DI resistivity internal, pressure relief, flow control, and pressure/flow control.
This unit is easy to use with intuitive interface and air and water filters that are changeable while unit is in operation.

Specifications :
Standard Chiller Process Fluid Temperature/Setpoint Range 5 °C to 40 °C
Ambient Temperature Range 10 °C to 40 °C
Temperature Stability ±0.1 °C
Cooling Capacity at 20 °C 900 W (3074 BTU)
Refrigerant R134A
Reservoir Volume 1.9 Gallons (7.2 Liters)
Foot print or Dimensions (H x W x D) 27.3 x 14.2 x 24.6 in (69.2 x 36 x 62.4 cm)
Weight with Turbine Pump 130.5 lbs (59.2 kg)
Voltage Requirement 115V/60Hz
Flow Rate- Turbine Pump 9.5 L/min

Tests Performed on the Thermo Scientific Thermoflex 900 Recirculating Chiller
Cooling capacity test is performed at 20°C to confirm chiller functionality
Flow rate is tested to verify pump is meeting manufacturer specifications
Chiller is temperature cycled numerous times to ensure longevity of instrument
Low temperature test is performed to ensure chiller is functioning within temperature range

Preventative Maintenance Performed on the Thermo Scientific Thermoflex 900 Recirculating Chiller
Pump is Inspected for Faults and Signs of Disrepair
Motor is Inspected for Shorts and Corroded Wiring
Plumbing is Inspected for Leaks and Corrosion
Filters are Cleaned and Replaced if Needed
Tank is Inspected for Leaks and Cleaned
Internal Circuitry is Inspected for Faulty/Shorted Components and Replaced Accordingly
Line Cord is Checked for Signs of Fray or Damage and Replaced Accordingly
Pump is Lubricated as Needed