Thermo Scientific Accela Autosampler HPLC Autosampler
AMERICA North (USA-Canada-Mexico)
Optimized for high-throughput applications and sub two micron column particles
Maximum reproducibility with isothermal injection and separation
Greater application flexibility with multiple sample formats
Superior robustness with a diamond coated high pressure valve Accela Photo Diode Array Detector
Highest detection sensitivity with proprietary LightPipe technology
Designed for U-HPLC separations with a 1cm path length, 2µL LightPipe flowcell
Maximum peak resolution with fiberoptic beam shaper
Easy maintenance with integrated wavelength validation
Specifications:
Sample capacity:
200 vials in five removable trays
288 samples in three 96-well microplates
1152 samples in three 384-well microplates
Standard vial:8 mL volume
Injection precision: < 1.0% RSD at 5 μL and up
Partial loop injection volume:
10 μL with standard 25 μL sample loop and 250 μL syringe
Up to 1000 μL injection volume with optional loops and syringes
Fixed loop (full loop) injection volume:
25 μL with standard loop and 250 μL syringe
Up to 1000 μL injection volume with optional loops and syringes
Sample carryover:
< 0.1%
6 mL flush volumes, user-specified for internal and external needle wash cycles
Minimum sample volume: 1 μL can be injected from 5 μL
Needle height control:
Programmable in 0.1 mm increments
Active vial bottom search selectable on/off
Column oven: Programmable in 1 °C increments from 5 to 95 °C
Tray temperature control:
Programmable in 1 °C increments from 0 to 60 °C
Typical vial temperature from 4 to 50 °C
Reagent vials: Four 16 mL vials can be specified for flush/wash or dilution functions
Remote inputs: Pump ready, inject hold
Remote outputs: Pump stop, autosampler ready, inject, gradient start, four timed events
Ambient environment:10 to 40 °C, 5 to 95% relative humidity, non-condensing
Power requirements: 100/120/220/240 Vac nominal; 50/60 Hz, 550 VA