Waveform: Truncated Exponential Biphasic.Waveform parameters adjusted
as a function of each patient’s impedance.
Energy: Adult - Nominal 150 Joules into a 50 ohm load. Infant/Child
- Nominal 50 Joules into a 50 ohm load. Automatically set based on type
of SMART Pads cartridge installed.
Shock-to-Shock Cycle Time: Typically less than 20 seconds between shocks
in a series
Quick Shock: Able to deliver a shock after the end of a CPR interval,
typically in eight seconds
Voice Instructions: Detailed voice messages guide responder through use
of the defibrillator
CPR Coaching: Instructions for adult and infant/child CPR available
at user’s option
Shock Delivery: Via adhesive pads placed on patient’s bare skin
as illustrated on pads
Controls: Green SMART Pads cartridge handle, green On/Off button,
blue i-button,orange Shock button
Indicators: Ready light; blue i-button; caution light
Dimensions:
Depth: 7.4 in/19 cm
Width: 8.3 in/21 cm
Height: 2.8 in/7 cm
Weight: 3.3 lbs/1.5 kg, with battery and pads case
Environmental:
Sealing: Solid objects per EN60529 class IP2X; Drip-proof per EN60529 class IPX1
Temperature: Operating - 32º - 122º F (0º- 50º C); Standby - 50º - 109º F (10º- 43º C)
Humidity: Operating - 0% to 95% relative,non-condensing; Standby - 0% to 75% relative,non-condensing
Altitude: Operating - 0 to 15,000 feet; Standby - 0 to 8,500 feet > 48 hours and 8,500 to 15,000 feet < 48 hours
Shock/Drop Abuse: Withstands 1 meter drop to any edge, corner or surface
Vibration Meets EN1789 random and swept sine, road ambulance specification
in operating and standby states
EMI (Radiated/Immunity) Meets EN55011 Group 1 Level B Class B and EN61000-4-3
Automated and User-Activated Self-Tests:
Daily Automatic Self-Tests: Tests internal circuitry, waveform delivery system,
pads cartridge and battery capacity
Pads Integrity Test: Specifically tests readiness-for-use of pads (gel moisture)
Battery Insertion Test: Upon battery insertion, extensive automatic self-tests
and user-interactive test check device readiness
Status Indicator: Blinking green “Ready” light indicates ready for use.
Audible “chirp” indicates need for maintenance
Data Recording and Transmission:
Infrared: Wireless transmission of event data to a PC or Palm® PDA,
using the IrDA protocol
Data Stored: First 15 minutes of ECG and the entire incident’s events
and analysis decisions