Perkin Elmer VICTOR X4 Multimode Plate Reader with stackers
AMERICA North (USA-Canada-Mexico)
Light sources: Tungsten-halogen lamp
– For fluorescence and absorbance measurements.
Xenon flash lamp
– For TR-fluorescence and UV absorbance measurements.
Detection units: Photomultiplier tube
– For fluorescence, luminescence and TR-fluorescence measurements.
Photodiode
– For absorbance measurements.
Plate formats: 1 to 1536
Shaking: Linear, orbital and double orbital. Duration, speed and amplitude are adjustable.
Temperature Control:
Temperature range: from 2 °C above ambient up to 50 °C (0.1 °C increments)
Uniformity: ±1 °C
Heating time: 10 minutes (from RT up to 37 °C)
Dispenser: The dispenser module consists of 1-4 high quality syringes. Depending on the dispenser option, up to 4 separate channels can be directed to one well (96-well plate).
Dispense volume: 5-350 µL (1 µL increments)
Accuracy typically: 5% for 5 µL, 0.5% for 50 µL, 0.05% for 350 µL
Precision typically: 1.4% for 5 µL, 0.2% for 50 µL, 0.02% for 350 µL
Speed: Minimum and maximum speeds are related to the volume.
Dead volume: 0.5 mL
Weight 155 lbs
Dimensions 26 × 29 × 22 in
Includes Power Cord
Voltage 120V/220V
hertz 50Hz/60Hz