Norland Excell
AMERICA North (USA-Canada-Mexico)
Features and Benefits
6' DXA table - using pencil beam technology
Standard sites: AP spine and proximal femur
Small footprint ideal for any office or mobile facility
Faster scan times:
AP Spine < 1.5 minutes Hip < 2.0 minutes
Exclusive Dynamic Filtration compensates for the widest range
in patient thicknesses to ensure every data point receives
proper exposure while eliminating photon saturation or
starvation Exclusive 77-step calibration process allows scanning
of the widest range of subjects without special calibration or
correction of data Norland-pioneered fracture risk assessment
feature provides simplified color reporting for physician and
patient Presents standard BMD (sBMD), using NHANES III data
a new standard for total spine and
total hip bone densitometry Technical Specifications
System
Scanner Description Pencil beam DXA
(dual energy x-ray absorptiometer)
Drive System Stepper motors, with Kevlar reinforced drive belts
X-ray Source Stationary anode x-ray tube,
air-cooling; 100kV constant potential, 1.3 mA anode current
(1.0 mA Japan), focal spot size: 0.5 mm
Filtration 8-level automated samarium filtering
(K-edge = 46.8 keV)
minimum filtration is 2.7mm aluminum equivalent X-ray
Detector Two Nal scintillation detectors in pulse counting mode
Power 100, 120, 220, 240 VAC +/- 10%, 50/60 Hz.,
700 VA maximum Operating Temp. 60-104 F, 15-40 C Relative Humidity Up to 80% non-condensing
Footprint
Dimensions Recommended dedicated floorspace: 84"L x 84"W (2134mm x 2134mm)
Tabletop size: 72"L [1828 mm] x 35.5"W [902 mm] at
26"H [660 mm]
Overall size: 72"L [1828 mm] x 48"W [1220 mm] x 49"H [1245mm]
Table surface to arm (maximum patient thickness): 22"L [559mm]
Weight 400 lbs. [181 kg]
Performance
Scan Sites AP spine, hip Scan Time