ICP DS 100, Sterilizing Oven
EUROPE (Western and Northern)
Brand : ICP
The dry sterilizer oven DS 100 is used for the sterilization
and depyrogenization of empty glassware or other glass
or stainless steel packaging or containment material
so that it can be introduced into a sterile room.
After the sterilization cycle, the material is therefore discharged
from the passage door of the sterile rooms.
Being the filters placed at the exit of the sterilization Chamber,
of the absolute type filters with cold efficiency of 99.998 DOP method,
during sterilization, the air circulating inside the trolleys
is air in class 100.
The filters used for this purpose are filters
that continuously withstand a temperature of 240 degrees centigrade
and therefore allow both dry sterilization and depyrogenation.
During all stages of treatment, inside the furnace,
the pressure is constantly maintained positive
with respect to the external environment.
After the automatic sterilization cycle,
the shutters all close to isolate the oven from the outside
while the ventilation continues to work to maintain
the internal overpressure ensuring that the material is not contaminated.
Sterilization time: 60 min for temperature 220 C while 75 min
for temperature 200 C.
Cooling time to be set around 120 min.
See for the use of timers the attached instructions
taking into account to always set the MODE "A" and the value in MIN.
The cooling time should be selected according
to the experience resulting from the type of load used
and the temperature of the sucked air.