Menu

Hitachi Airis Elite 0.3T

Ref : 2528100-14-CP
Condition : Used
Manufacturer : Hitachi
Model : Airis Elite 0.3T
Year(s) : -
Quantity : 1
Location : Seller or machines location:
AMERICA North (USA-Canada-Mexico)

AIRIS Elite 0.3T is a combination of the renowned reliability, cost-effectiveness
and patient comfort of the AIRIS II with throughput
and capabilities previously not seen at Open low-field.
The Elite’s permanent magnet design incorporates
an asymmetric two-post architecture, for maximum openness.
The proven performance of AIRIS Elite’s vertical field permanent magnet
with higher-order active shim technology, Scalable DualQuad RF system,
and broad sequence abilities are all integrated through Hitachi’s unique approach
to optimize system performance.

AIRIS Elite continued the tradition of Hitachi's leadership in Open MR technology,
providing new capabilities to low-field MRI including parallel imaging
and RF fat saturation.

Features:
Higher Order Active Shim enables RF fat saturation at 0.3T
21mT/m and 55T/m/s gradients for high resolution and fast scan time
RAPID parallel imaging for shorter scan times
Gantry opening: 43 cm
Table top width: 80 cm
Vertical movement
Power-driven
Range 18 – 32 cm
Lateral movement
Power-driven
Range ±4 cm
Longitudinal travel speed
Power-driven
4.7 cm/sec maximum
Positioning
Three-plane
Laser light localization
Positioning accuracy: ±1mm
Emergency evacuation: Manual table-top release
Weight limit: 500 lbs
Communication
Two-way intercom
Technologist alert system

Magnet System:
Permanent magnet type
Vertical magnetic field
0.3T field strength
Homogeneity: 0.4ppm@20cm DSV (FWHM)

Shielding:
Self-shielded
Passive

Shimming:
Computer-aided passive shim
Gradient shim
Active shim (4 channels plus B0 compensation)

5G fringe field:
6.6 ft (horz.) x 8.2 ft (vert.)

Other machines similar to Hitachi Airis Elite 0.3T

1
Location : ASIA (North East)
Year(s) : 2015
1
Location : ASIA (North East)
Year(s) : 2014
1
Location : ASIA (North East)
1
Location : ASIA (North East)
Year(s) : 0