GE, OEC 9600 C-Arm
AMERICA North (USA-Canada-Mexico)
Generator:
High frequency
4.0 kW full-wave
Up to 120 kVp
Up to 75 mA for radiographic film exposures
Continuous fluoro boost mode up to 12 mA (optional)
Pulsed fluoro mode
Pulsed fluorography mode up to 40 mA (optional)
Full power from outlet:
– 50 Hz or 60 Hz
– 100V-15A to 240V-8A
Patented energy buffer design
X-Ray Tube:
Rotating anode X-ray tube
Focal spots: 0.3 mm and 0.6 mm
Anode heat capacity: 300,000 HU
Anode cooling rate: 60,000 HU/min
Housing heat capacity: 1,600,000 HU
Anode diameter: 3.1 in.
Anode angle:
– 8.5 in. (9-inch Image Intensifier)
– 10 in. (12-inch Image Intensifier)
X-ray tube cooling fan (Vascular, Neurovascular & Cardiac units)
Digital Image Rotation:
Digitally adjusts image display
Image update
-Image rotation
-Image reversal (side-to-side)
-Image invert (top-to-bottom)
Image positioning without additional exposure
Collimator:
Iris collimation with dual opposing semi-transparent shutters allow elliptical, round and rectangular collimation
Iris and shutters are continuously variable
All functions remotely controlled from the C-Arm control panel
Fluoroscopy Mode:
Focal spot: 0.3 mm
kVp range: 40 – 120 kVp
mA range: 0.2 – 5.0 mA normal mode
mA range: 1.0 – 12 mA continuous fluoro boost mode
Auto and manual modes
Continuous, one-shot or pulsed operation
ABS varies mA, kVp and came
Pulsed Fluoro Mode (optional):
Standard with the Expanded Surgical Platform (ESP) and higher configurations:
Focal spot: 0.3 mm
kVp range: 40 – 120 kVp
mA range: 0.2 – 40.0 mA
Pulse rate: 1, 2, 4, or 8 pulses per second
Pulse width: 30 or 50 milliseconds
Computer controlled ABS, mA, kVp and camera gain
Digital Spot Mode (optional):
Standard with the Expanded Surgical Platform (ESP) and higher configurations:
Enhanced digital fluorographic one shot with 12 mA boost
Automatically terminates exposure and stores enhanced image to storage archival disk
Radiographic Mode:
Focal spot: 0.3 mm or 0.6 mm
Focal spot automatically selected
0.3 mm – mAs range: 1 – 100 mAs
0.6 mm – mAs range: 110
Video Imaging System:
9″ Image Intensifier:
Tri-mode 9/ 6/ 4.5 in. image intensifier
Minimum central resolution (at the monitor):
– 9 in.: 1.4 lp/mm
– 6 in.: 2.0 lp/mm
– 4.5 in.: 2.4 lp/mm
DQE: 65% (typical)
Video Camera:
High resolution CCD camera
Full frame capture
360° motorized rotation
On-screen orientation indicator (realtime feedback without fluoro)
Left-right image reversal
Top-bottom image reversal
Negative mode
Video signal: Standard RS 170A 60 Hz, 525 lines
Aspect ratio: 4:3
Computer controlled features:
– Gain
– Blanking
– Camera iris
Video Display (OEM Factory):
Dual 16 in square, black etched, anti-glare, anti-static
525 lines/frame – 512 pixels
Ambient room light sensor
Multi-LED X-ray “ON” indicator
Remote window/level controls
Video Display (Upgraded Retrofit Kit):
32″ LED Lit LCD Display With Dual Screen Capabilities
Anti-glare, progressive full frame scan monitor
2 Mega Pixel high-resolution output
Ambient room-light compensation
C-Arm Dimensions:
Height: 68 / 86 in (1734 / 2191 mm)
Width: 33 in (838 mm)
Length: 78 in (1994 mm)
Weight: 720 lbs.
Workstation Dimensions:
Height: 64 in (1632 mm)
Width: 27 in (692 mm)
Depth: 27 in (692 mm)
Weight: 430 lbs.