GE Aespire S5 Anesthesia Machine with 7900 Smartvent
AMERICA North (USA-Canada-Mexico)
Features
Enhanced monitor integration capabilities with
the GE / Datex Ohmeda Anesthesia Monitor Series.
Lightweight and compact for easy maneuverability
Optional integrated auxiliary O2 flowmeter and
suction control
Advanced Breathing System (ABS™)
One step bag/vent switch turns the ventilator on/off
Minimal number of parts and
tube connections greatly reduces the potential for
leaks and misconnects
Ease of disassembly (no tools)
Fully autoclavable and latex-free
7900 SmartVent Ventilator
Volume and Pressure modes with electronic PEEP
Synced intermittent mandatory ventilation ( SIMV )
Direct access to ventilator parameter settings
PSVPro ( Pressure support with Apnea backup mode. )
Smart alarms direct user to specific problems and
affected parameters
Ideal for spontaneously breathing patients
Improved low flow/reduced life cycle costs
Only one scheduled maintenance check per year
Fresh gas flow compensation –
automatically (available with
tidal volume compensation option)
Minimum O2 flow of 50 mL
Dual air flow tubes standard for higher resolution of
low flows
SPECIFICATIONS
Physical Weight & Dimensions
Height: 134.5 cm/52.9 in
Width: 72 cm/28.3 in
Depth: 73 cm/28.7 in
Weight: Approximately 108 kg/238 lb
Top Shelf
Weight limit: 34 kg/75 lb
Width: 66 cm/26 in
Depth: 40 cm/15.75 in
Work Surface
Height: 81.7 cm/32.2 in
Size: 2160 cm-2/334 in-2
DIN Rail
Side of machine: 34.5 cm/13.6 in
Drawers (internal dimensions)
Height: 17.5 cm/6.9 in
Width: 33 cm/13 in
Depth: 26.5 cm/10.4 in
Absorber Bag Arm (optional)
Arm length: 30.5 cm/12 in
Bag arm height (adjustable): 87 cm (34.3 i) to 104 cm (40.9 in)
Casters
Diameter: 12.5 cm/5 in
Brakes: Individual locking front casters
Electrical Specifications
Current Leakage
100/120 V: < 300μA
Power
Power input: 100-120 Vac, 50/60 Hz
Power cord: Length: 5 m/16.4 ft
Rating: 15A @ 120 Vac
Inlet/outlet modules (120 V)
System circuit breakers: 15A
Outlets (optional): 4 outlets on back, 3-2A, 1-3A individual breakers, optional isolation transformer
Pneumatic Specifications
Auxiliary Common Gas Outlet
Connector: ISO 22 mm OD and 15 mm ID
Gas Supply
Pipeline input range: 240 kPa to 600 kPa/35 psig to 88 psig
Pipeline connections: DISS-male, DISS-female,
DIN 13252, AS4059, S90-116 or NIST (ISO 5359).
All fittings available for O2, N2O, and Air, and contain pipeline filter and check valve.
Cylinder input: Pin indexed in accordance with CGA-V-1 or DIN (nut and gland); contains input filter and check valve
Note: Maximum 3 cylinders; two inboard mounted, one outboard mounted.
Primary regulator diaphragm minimum burst pressure: 2758 kPa/400 psig
Primary regulator nominal output: < 338 kPa/49 psig Pin indexed cylinder connections
< 407 kPa/59 psig DIN cylinder connections
Environmental Specifications
System Operation
Temperature: 10° to 40°C/50° to 104°F
Humidity: 15 to 95% relative humidity (non-condensing) per IEC 68-2-3
Altitude: -440 to 3565 m/500 to 800 mmHg
System Storage
Temperature: -15° to 50°C/-5° to 122°F
Humidity: 10 to 95% relative humidity (including condensing) per IEC 68-2-3
Altitude: -440 to 3565 m/500 to 800 mmHg
Oxygen cell storage: -15° to 50°C/5° to 122°F; 10 to 95% relative humidity; 500 to 800 mmHg
Electromagnetic Compatibility
Immunity: Complies with all requirements of EN 60601-1-2
Emissions: CISPR 11 group 1 class B
Approvals: UL 2601-1; CSA C22.2 #601.1; EN/IEC 60601-1; CE 0197