Drager Narkomed MRI-2 Anesthesia Machine
AMERICA North (USA-Canada-Mexico)
Can be used in shielded or unshielded MRI environments of up to 3.0 Tesla.
Integrated monitor measures and displays airway pressure, volume and oxygen data with airway pressure waveform.
Customizable alarms for each unique situation.
Standard Narkomed patient suction system with regular or reusable canister designed for MRI use.
3-hour battery backup guarantees continuous operation of all machine functions.
Accurate ultrasonic flow sensor has no moving parts for easy cleaning.
Time-cycled, pressure-limited AV2+ ventilator equipped with pneumatic circuitry and independent breathing rate and tidal volume controls.
Vapor 2000 needs no annual recalibration, compensates for variances in pressure, temperature and flow.
Open or closed waste scavenging system.
Dual canister absorber for increased time between absorbent changes.
Compatible with halothane, isoflurane, enflurane and sevoflurane anesthetic agents.
SPECIFICATIONS
General
Dimensions: 31.5” (80 cm) W x 53” (135.3) H x 27” (68.6 cm) D
Weight: 230 lbs (104.3 kg)
Weight of Battery Charger: 3 lbs (1.36 kg)
Electricity
120 Volt Power Supply: 100 - 120 VAC @ 50/60Hz
220/240 Volt Power Supply: 200 - 240 VAC @ 50/50Hz
Battery
Reserve Power Time: ≤ 3 hours
Charging Time: ≤ 16 hours
Gas Delivery System
Number of Gases: 3 gas capability (O2, N2O, Air)
Oxygen Flush Flow Rate: 55 L/min
Low Oxygen Supply Pressure Alarm: 30 - 40 psi
Vaporizer
Vapor 2000
Positions: 2
Mounting: Fixed
Spec Title
AV2+
Frequency: 1 - 99 (in 1 BPM increments)
I:E Ratio: 4:1 -1: 4.5
Inspiratory Flow: 10 - 100 L/min
Tidal Volume: 20 - 1500 mL
Pressure Limit Adjustment: 15 - 120 cmH2O
Oxygen Monitoring
Display Range: 10 - 100%
Resolution: 1 vol% O2
Airway Pressure Monitoring
Display Range: -10 to 125 cmH2O
Resolution: 1 cmH2O
Tidal Volume Monitoring
Display Range: 0.1 - 2.0 L
Resolution: 0.01 mL
Utility Requirements
System operates with 120 VAC @ 50/60 Hz electrical source.
Regulator assemblies and yokes for
external O2 and N2O cylinders.
O2, N2O and Air hoses for connection pipeline sources to
DISS machine connections.