Datascope CS300
EUROPE (Western and Northern)
The Maquet Datascope CS300 IABP with
IntelliSense combines fiber-optic speed with
automatic in vivo calibration.
The result is a faster time for therapy,
faster signal acquisition, and faster adaptation to
rate and rhythm changes.
Maquet has the only fiber-optic IABP and catheter system that
automatically calibrates in the patient after insertion and
automatically recalibrates in vivo every two hours or
sooner should the patient or environmental conditions change.
The Datascope CS300 IABP with IntelliSense software is
designed to emulate an experienced user,
allowing you more time to focus on your patient.
This sophisticated algorithm automatically selects the best lead
and trigger source, sets optimal inflation and
deflation timing, and still allows the clinician the ability to
fine-tune deflation timing.
The Maquet IABP quickly adapts to rate or rhythm changes in
order to optimize augmentation and support during diastole.
The CS300 with intelliSense is designed to emulate an
experienced user, allowing the clinician more time to
focus on the patient. The CS300 regularly examines key
landmarks on the ECG and arterial pressure waveform to
determine optimal timing.
It then quickly adapts to rate or rhythm changes instead of
imposing fixed timing rules.
Features:
Faster inflation and deflation speed
CS300 has true one-button start-up
Automatically calibrates the fiber-optic pressure sensor in
the patient and recalibrates every 2 hours or
sooner should patient or environmental conditions change
Automatically evaluates and selects the best lead and
trigger source
Automatically sets optimal timing
Automatically adjusts to changes in
patient conditions without clinician intervention
A Maquet fiber-optic IAB catheter
Conventional fluid-filled IAB catheters
Trigger
ECG Trigger: Threshold dynamically adjusted by
the system for high sensitivity and selectivity of
the R-wave detection;
Minimum = 120µV ± 20µV at a normal gain; 40µV at max. gain
Pressure Trigger: The default trigger threshold is
automatically adjusted to 38% of the systolic pulse height;
7 mm minimum
Manual threshold mode: User-adjustable between 7 and 30 mmHg ± 3 mmHg
Pacer A Trigger: R-wave detection (as above
) except pacer blanking is extended to 100 ms
Pacer V/A-V Trigger: V Pacer: fixed at a rate up to 185 bpm
(no demand pacing)
A-V Pacer: fixed at a rate up to 125 bpm (no demand pacing) with A-V intervals between 80-224 ms
ECG
ECG Leads: In Auto Operation Mode: I, II, III,
External In Semi-Auto Mode: I, II, III, AVR, AVL, AVF, V,
External (12 lead compatibility)
ECG Gain (default): 1 V output per 1 mV input ±5%
(waveform automatically scaled to occupy ECG display window)
Gain (variable): 0.15 to 3.0 cm/mV ±20% (autoscaling disabled)
Frequency Response: 0.5-12 Hz (display);
0.5-135 Hz (Output to External Monitor)
Defibrillator Protection: Discharge level = 360 J
(trace returns to the screen in 5-sec max)
ESIS: Automatic suppression with internal ECG amplifier