Criticare 8100 Poet Plus
AMERICA North (USA-Canada-Mexico)
8100 Poet Plus Vital Signs CO2 Veterinary Monitor - Preowned in Good Condition
VENDOR DESCRIPTION
Poet Plus® 8100 is the perfect choice for busy departments demanding maximum flexibility, unsurpassed ease of use, and accurate, reliable performance. Respond to every monitoring need, from standard vital signs to conscious sedation. Monitor adults, pediatrics, and neonates in virtually every department, at bedside or in transit. The Poet Plus 8100 mounts quickly and easily to walls, roll stands, and bedrails.
With the compact, lightweight Poet Plus 8100, you focus on your patients, not your monitor. The innovative user interface with simple, one-level menus and dedicated function keys ensures super-fast set-up. The big, bright display with large numerics is easily seen, even from across the room.
FEATURES
Easy-to-read 10.4" active TFT color display
View up to 6 waveforms simultaneously.
Easy to learn, easy to use menus, dedicated function keys, and simplified controls
Smart battery management with rechargeable "hot-swappable" sealed lead acid batteries
Lightweight design and versatile mounting options
Adapts readily to department and transport requirements
SPECIFICATIONS
System Outputs
Communication Ports: RS-232 compatible; DB9, analog mini-din
Nurse Call: contact switch, stereo jack 3.5 mm, length 14 mm standard
Defib / Sync: BNC connector
Video Port: high density D-sub 15-pin
Recorder
Type: Thermal Line Printer
Paper speed: 12.5, 25, 50mm/sec
Mechanical
Weight: Base unit without batteries: 12 lbs, 5.45kg
Size: Base unit: 13.25 "(H) x 12.5" (W) x 8.75 "(D)
Power Requirements
Voltage: 100, 120, 220, 240 VAC, 50/60 Hz
Consumption: 35 W, typical
Batteries: 2, Sealed Lead Acid, life: 2 hr, typical recharge time: 4.5 hours
ECG
Connectors: 3- or 5-lead
Leads: User-selectable; I, II, III, aVR, aVL, aVF, V (5-lead); I, II, or III (3-lead)
Gain Selection: 1/2mV, 2mV, 4mV
Frequency Response:
- Diagnostic: 0.05 to 100 Hz (-3db);
- Monitor: 0.5 to 40 Hz (-3db)
Sweep Speed: 12.5, 25, 50mm/sec
Electrosurgery Protection: Yes
Defibrillator Protection: Yes
Pacer Detection/Rejection: Yes
Non-Invasive Blood Pressure
Technique: Oscillometric measure upon inflation
Automatic Measurement Cycles: 2, 3, 5, 10, 15, 30 min; 1, 2, 4 hr
Measurement Range: Adult: 30 to 300mmHg; Neonate: 0 to 150 mmHg
Resolution: 1mmHg
Transducer Accuracy: ±2mmHg or ±2% of reading, which ever is greater, over full range
STAT mode: 5 min of consecutive readings
Pulse Oximetry
Range: 1% to 99%
Resolution: 1%
Accuracy: 70% to 99% range: ±2%; 50% to 69% range: ±3%;