Amersham, Biosciences AKTA FPLC Explorer 100 System
AMERICA North (USA-Canada-Mexico)
UV-900 Detector Specifications :
Wavelength range :
190–700 nm in steps of 1 nm,
3 wavelengths simultaneously
Bandwidth 4 nm
Wavelength accuracy ±2 nm
Wavelength reproducibility
< 500 ms (one cycle from 214 nm to
254 nm and back to 214 nm)
Linearity < 2% deviation up to 2 AU at 260 nm with
Uracil at pH 2
Static noise
short term: < 6x10-5 AU at 230 nm
long term: < 6x10-5 AU at 230 nm
Drift < 2x10-4 AU/h at 254 nm
Flow Cell :
Optical path length 2 mm cell 2 mm, 2 µl cell volume
Optical path length 10 mm cell 10 mm, 8 µl cell volume
Max. flow rate 100 ml/min
Max. pressure 2 MPa (290 psi)
Temperature range +4 to +40 °C (liquid)
Wetted material:
PTFE (polytetrafluoroethylene)
PEEK (polyetheretherketone)
titanium (palladium alloy),
quartz (synthetic fused silica) and titanium
P-900 Pump Specifications
Flow Rate Range :
Isocratic mode: 0.01 to 100 ml/min in steps of 10 ul/min
Gradient mode: 0.01 to 100 ml/min in steps of 10 ul/min
Double mode: 0.01 to 200 ml/min in steps of 10 ul/min
Pressure Range 0 to 10 MPa (100 bar, 1450 psi)
pH Stability Range 1 to 13 (1 to 14 with