Alaris Medley 8110
AMERICA North (USA-Canada-Mexico)
Dimensions: 4.5″W x 15.0″H x 7.5″D
Weight: 4.5 lbs
Operating Principle: Positive Displacement
Flow Rate:
0.01-9.99 ml/hr in 0.01 ml increments
10-99.9 ml/hr in 0.1 ml increments
100-999 ml/hr in 1 ml increments
Volume to be Infused:
0.1-9.9 ml in 0.01 ml increments
10-60 ml in 0.1 ml increments
Rate Restriction by Syringe Size:
50-60 ml syringe – 0.1-999 ml/hr
30 ml syringe – 0.1-650 ml/hr
20 ml syringe – 0.1-500 ml/hr
10 ml syringe – 0.1-250 ml/hr
5 ml syringe – 0.1-150 ml/hr
3 ml syringe – 0.01-100 ml/hr
1 ml syringe – 0.01-30 ml/hr
KVO Rate:
1ml/hr or selected rate, whichever is lower
Infusion Pressure Maximum:
800 mmHg without pressure sensing disc
1060 mmHg with pressure sensing disc
Rate Accuracy:
±2% of full scale plunger travel not including syringe variation