Alaris, Cardinal Imed Gemini PC-1 Infusion Pump
AMERICA North (USA-Canada-Mexico)
Product Specifications
Operating Principle:
Linear peristalsis
Dimensions:
10.8″ x 8.2″ x 6.5″ (27.5 cm x 20.6 cm x 16.5 cm)
Weight:
11.3 lbs (5.1 kg)
Power Requirements:
90 – 130 (120 nominal) VA, 0.5A, 50/60 Hz, 3 wire, single phase
Battery Operation:
With a new battery, approximately 5 hours At 125 ml/hr before a “HELP BATTERY”
Electrical Leakage:
Less than 100 microamps
Delivery Rate Range
Micro:
0.1-999 ml/hr (Note: Recommended maximum rate in the Controller Mode is 500 ml/hr
Macro:
0.1-99.9 ml/hr (in 0.1 ml/hr increments)
Volume to be Infused
Micro:
0.1-999.9 m/l (in 0.1 ml increments)
Macro:
1-9999 ml (in 1 ml increments)
KVO (Keep Vein Open) Rate:
1 ml/hr if set rate is 1 ml/hr or above, or actual rate if set at 0.9 ml/hr or below
Occlusion Pressure
Pump mode:
10 ±2 psi for rates above 30 ml/hr. For rates of 30 ml/hr and below, the occlusion pressure will be less than 10 psi, and is rate dependent (unless locked into 10 psi for all rates)
Advisories:
Secondary, Primary, Infusion Complete – KVO, Empty Container-KVO, Low Battery, Maintenance, Pause, On Batt, 10 psi. Flow, Total Vol Infused, Sec Vol Infused Macro, Micro, Versataper: Setup, Step: xx, Steps: xx, Total Vol: xx, Duration: xx, Versataper, and Time Left:xx
Prompts:
Select P/C mode, set rate, set VTBI, and Press Start
Alarms:
Check ECD, Check IV Set, Close Door, Air-in-Line, Occluded Patient Side, Occluded Fluid Side, Occluded, Partial Occlusion Fluid Side, Flo-Stop Open/Close Door, KVO Comm Cable, KVO Comm Tout, and KVO
Malfunctions:
Help internal error, help battery, and (Watchdog Malfunction)
Secondary (Piggyback):
Dual rate programming