Agilent Technologies G8481A Recirculating Water Chiller
AMERICA North (USA-Canada-Mexico)
The Agilent Technologies G8481A Recirculating Water Chiller for the Agilent 700 Series ICP Emission Spectrometer
can be used to ensure the delivery of coolant at the correct pressure and temperature.
The water chiller is equipped with a water flow sensor, which will stop operation of the plasma and camera
Peltier assembly if the cooling water flow through the instrument drops below 1.1 L/min (0.3 gal/min).
Pressure regulation is recommended for water chiller systems that may exceed
the maximum permissible pressure of 310 kPa (45 psi) or may be subject to pressure fluctuations.
The water chiller generates approximately 1,000 watts (joules per second), or 3,6000 kilojoules per hour (3,412 BTU per hour).
Specifications:
Electrical Requirements: 220-240VAC, 50Hz, 1450 VA
115VAC, 60Hz, 1645 VA
Voltage ±10%, Frequency ±1Hz
Cooling capacity:
Axial system: 1 kW
Radial system: 200W
Flow rate: 1.1 L/min (0.3 gpm) minimum
Recommended inlet temperature: 20°C (68°F)
Maximum for Agilent 720 Series ICP-OES: 30°C (86°F)
Maximum for Agilent 710 Series ICP-OES: 4°C (113°F)
Minimum inlet pressure: 55 kPa (8.0 psi)
Maximum inlet pressure: 310 kPa (45 psi)
Weight: 181 lb. (82 kg)
Dimensions (W x H x D): 14.5 x 22.6 x 27.6 in. (368 x 575 x 702 mm)