Waters 2465 Electrochemical Detector
The Waters 2465 Electrochemical Detector combines sensitivity, reliability and simplicity for HPLC electrochemical detection. Its versatile detection modes are ideally suited for analyzing a wide range of compounds.
The Waters 2465 Electrochemical Detector’s multiple flow-cell configuration combines working and reference electrode designs to minimize cost and enhance flexibility. The flow cell accommodates several reference electrodes: ISAAC; traditional salt-bridge design, which limits air bubble formation; and Hy-Ref, which is ideal for carbohydrate analysis and is compatible with extreme mobile phases.
The Waters 2465 Electrochemical Detector is designed for ultra-trace analysis in standard and microbore Alliance® HPLC based, Empower™ Software controlled applications.
Features:
• Multiple detection modes, including direct current, pulsed amperometric, and scanning
• Easier maintenance with front-mounted flow cell for simple cleaning and assembly
• Stabilized temperature for better accuracy with the integral flow cell and column heater compartment
• Faraday-shielded oven with thermal operating stability and reduced noise and drift characteristics
Condition Specification
Line frequency 50 to 60 Hz
Line voltage 100 to 240 VAC (self-selecting)
Operating modes DC, Pulse, and Scan
Potential range Between +2.00 and –2.00 V in 10-mV increments
Autozero Triggered by keyboard, rear panel contact closure, or RS-232C control
RS-232C Full control and data acquisition of all parameters
Detector oven Height 37 cm, temperatures from 7 °C above ambient to 45 °C, accuracy 0.5 °C, stability 0.1 °C; accommodates HPLC column and flow cell
Rear panel I/O connections • Main power
• A (15-pin Sub D) – Input: Cell on, cell off, Start,
Autozero, or Reset; Output: Relay 1, Relay 2, AUX 1, AUX 2, Overload
• B (15-pin Sub D) – Inject marker (output) and Baseline marker (input)
• C (phone jack) – Not applicable
• OUTPUT (BNC) – Analog signal
• RS-232C (9-pin Sub D) – Full instrument control; data acquisition
• Ground stud – Instrument ground connector
Operating temperature range 39 to 104 °F (4 to 40 °C)
Storage temperature range –104 to 158 °F (–40 to 70 °C)
Relative humidity range 20 to 80%, noncondensing
Storage humidity range 0 to 90%, noncondensing
Spacers 25, 50, or 120 µm
WE diameters 2.00 and 2.74 mm (0.75 mm with micro flow cell)
Cell volume 0.150 µL minimum with 3-mm flow cell
0.080 µL minimum with 2-mm (standard) flow cell
0.011 µL minimum with 0.75-mm (micro) flow cell
Working electrodes Glassy carbon (GC), gold, platinum, and silver
Reference electrodes Salt-bridge Ag/AgCl; ISAAC; Hy-REF
Auxiliary electrode Stainless steel
Wetted materials PCTFE, FEP, 316-SS (stainless steel), Viton®
, silver, silver
chloride, gold, and platinum
Pressure Maximum 40 psi (2.76 bar, 276 kPa)
Flow rate • Standard flow cell: 25 µL/min to 2.0 mL/min
• Micro flow cell: 1 to 25 µL/min
This refurbished Waters 2465 Electrochemical Detector is in great condition.