Varian GC CP 3800 without Detetors, Dual SSIs good for Varian Saturn MSD

Dimensions and Weights

Size: 53 cm (h) x 66 cm (w) x 56 cm (d)
Weight: 43 kg

Environmental Conditions

Operating temperatures: 10°C to 40°C
Operating humidity (relative): 5% to 95%
Line voltage requirements: 101V, 120V, 230V (±10% nominal)

Column Oven

Dimensions: 28 cm (w) x 20 cm (d) x 28 cm (h)
Temperature range: ambient +4 °C to 450 °C
With liquid N2: -99 °C to 450 °C
With liquid CO2: -65 °C to 450 °C
Temperature program ramps/holds: 7/8
Maximum temperature ramp rate: 100 °C/min for all voltages
Cool down rate: 400 °C to 50 °C in 4.5minutes
Temperature set-point resolution: 1 °C

Injector Options

Maximum injectors: three operating concurrently
Pneumatics: manual or electronic flow control (EFC)
Columns: capillary columns standard with optional packed column adapter
Injectors:
CP-1177Split/Splitless
1079 Programmable Temperature Vaporizing (PTV)
1061 Flash Megabore/Packed
1041 On-Column Megabore/Packed

CP-1177 Split/Splitless

Pressure range: 0-100 psi
Total flow: 500 rrillrnin at 10 psi
Maximum temperature: 450 °C
Split range: 1-10,000 (column dependent)

1079 Programmable Temperature Vaporizing (PTV)

Pressure range: 0-100 psi
Total flow: 500 mL/min at 10 psi
Temperature range
99 °C to 450°C liquid N2
65 °C to 450°C liquid CO2
Maximum temperature ramp rate: 180 °C/min
Temperature ramps: 4
Split range: 1-10,000 (column dependent)
Operational modes
Temperature ramped splitless
Coldon-column
Large Volurne Injection
Split and splitless
ChromatoProbe solid sample introduction: optional

1061 Flash Megabore/Packed
Pressure range: 0-100 psi
Total Flow: 50 mL/min (Type 3 EFC)
500mL/min (Type 4 EFC)
Maximum temperature: 450 °C

1041 On-Column Megabore/Packed
Pressure range: 0-100 psi
Total Flow: 50 mL/min (Type 3 EFC)
500mL/min (Type 4 EFC)
Maximum temperature: 450 °C